Fluke Biomedical ASM-990 Series User Manual
Page 19

Operation
Basic Operation
2
2-11
Date/Time Adjustment
When the “Date/Time” adjustment mode is entered, the display will appear as in Figure 2-24. The “MM”
will be highlighted first. If the user takes no action in five seconds, the “DD” will then be highlighted.
Similarly, the “YY”, “HH” and “MM” selections will be highlighted individually, with a five second interval
between each. When an item is highlighted, the user can modify its value by pressing the “UP” or
“DOWN” arrow key. If there is no action from the user in five seconds after releasing the “UP” or “DOWN”
arrow key, the highlighted area will move to next item. The “AM” or “PM” will be set automatically
according to the “HH” setting.
Press the “SEL” key to save the displayed value and return to the “NORMAL OPERATING SCREEN.”
(Note: The “DATE/TIME” will not require frequent adjustment since the internal real time clock continues
to operate when the unit is off.)
CAUTION
Only lithium battery manufactured by Sanyo Model
CR1220 rated at 3.0 volts DC is to be used.
DATE/TIME
MM/DD/YY HH:MM AM
Figure 2-24.
Date/Time Setup
Alarm Settings
Alarm settings may be reviewed/modified by first selecting “ALARM” from the “SETUP SUB-MENU,” then
pressing “SEL.” The screen will now appear as in Figure 2-25. To review the alarm settings only, use the
“UP” or “DOWN” arrow key to highlight “REVIEW SETTINGS;” then press “SEL.” The screen will now
appear as in Figure 2-26.
VIEW ALARM SETTING
cps
60.00
ALARM SUB-MENU
REVIEW SETTINGS
MODIFY SETTINGS
ENABLE ALARM
Figure 2-25. Alarm Sub-Menu
Figure 2-26.
View Alarm Setting