2 specifications – Fluke Biomedical 37-720 User Manual
Page 6

37-720
User Manual
1-2
1.2 Specifications
Range
cGy: 0 to 2000
cGy/min: 0 to 2000
Sensitivity
.25 nC/cGy – 2 nC/cGy
Calibration Adjustments Front panel screwdriver adjustments:
CAL A and CAL B used for calibration of detector A and detector B.
ZERO A and ZERO B used to set up zero for detector A and detector B. CAL
A and CAL B used for calibration of detector A and detector B.
ZERO A and ZERO B used to set up zero for detector A and detector B.
Front Panel Controls
Power Switch: ON/OFF
Function Switch: DOSE or RATE modes of operation
Detector Switch: DET A or DET B selects detector input.
Reset Switch: Resets & zeros monitor for dose measurements.
Rear Panel Controls
Two BNC connectors for DET A input and DET B input.
Display
A 3 ½-digit liquid crystal display provides direct reading of the measurement
Accuracy
+ 5 % of full scale, current and charge
Rads Holding Drift
Better than 1 % / minute
Batteries Required
One 9 V alkaline cell (Duracell MN1604 or equivalent)
Battery Check
LO BAT appears on display when battery voltage drops below approximately 6 V
Battery Life
Approximately 150 hours
Case
Ivory, high impact plastic
Dimensions
2.5 in x 6.25 in x 8 in
(H x W x D)
(6.4 cm x 15.9 cm x 20.4 cm)
Weight
1.5 lbs (0.68 kg)
Operating Conditions
10 °C to 45 °C (50 °F to 104 °F)
Maximum 90% Relative Humidity non-condensing