Installation
Mounting
3
3-1
Section 3
Installation
 
3.1 Mounting 
The mounting bracket supplied may be attached to any convenient vertical mounting surface. Before 
mounting the bracket to the wall, however, make sure that the six-foot converter power cord will reach the 
nearest power outlet. 
The wall-mounting bracket comes supplied with double-sided foam tape already attached. Just peel off 
the protective paper covering and firmly apply to the mounting surface. 
 
 
 
Once in place, the bracket cannot be moved or 
removed without destroying the tape. Make sure of 
your installation site before applying. If you wish to 
attach the bracket with screws, mounting holes are 
also provided. 
3.2 Power
The plug on the end of the converter power cord should first be inserted into the jack on the bottom of the 
Primalert. The converter should then be plugged into the wall outlet. 
3.3 Optional Power
The Primapak Backup battery pack (P/N 05-441) provideds a sourced of 12 to 16 VDC power.
 
 
 
The converter plug supplies this voltage.
 
The plug is a Switchcraft S760, or equivalent, with the sleeve negative and pin positive. The current 
drain at background radiation levels is about 25 mA and about 250 mA at full alarm levels. 
3.4 Grounding
If it is desired to ground the instrument with a separate ground wire, as may be required under certain 
codes, Figure 3-1 shows how to accomplish this. Simply remove the back cover by loosening the single 
screw in the center of the cover. Note the position of the ground terminal in Figure 3-1. Attach the wire by 
running it up through the small hole between the jacks. At least a #16 gauge wire should be used for this 
purpose. 