Section 1, Introduction, 1 general description – Fluke Biomedical 960SG-500 User Manual
Page 5: 2 application, 3 specifications
1
Introduction
1-1
Section 1
Introduction
1.1 General Description
960SG Signal Generator is utilized to verify the proper operation of the Universal Digital Rate meter 942 Monitoring
Systems and the Universal Digital Rate meter 956 Monitoring Systems. It simulates pulses from the radiation detector
and is applied to the radiation pulse counting circuit. The 960SG can be supplied as part of a radiation monitoring
system or as a separate, stand-alone piece of test equipment. The Model 960SG-500 is designed to mount in a 3-bay
rack chassis. Figure 1-1 is the Signal Generator functional block diagram.
The Model 960SG-500 utilizes two independent 960SG boards in a UDR style housing, for UDR applications. Front
panel controls include decade range, variable amplitude and variable frequency.
1.2 Application
The signal generator provides pulses to the UDR through the signal input jack. The user can select, via a slide switch,
each decade in CPM from 0 CPM to 1 x 10
5
CPM. A variable frequency control is located on the front panel to allow
adjustment within the selected range.
1.3 Specifications
The specifications for the signal generator are listed below:
Switch Selectable Frequency
Counts Per Minute:
FREQUENCY CPM
0 0
.167
Hz
10
1.67
Hz
100
16.7
Hz
1
k
167
Hz
10
k
1.67
kHz
100
k
Output Pulse (BNC Connector)
Positive polarity, 2 V pulse height, 800 ns duration
Output Impedance
51 ohms
Dimensions (H x W x D)
3.5 in. x 5.6 in. x 13.5 in.
(8.9 cm x 14.2 cm x 34.3 cm)
Weight
Approximately 2 lbs (0.9 kg)
Operating
Temperature
32
°F to 122°F (0°C to 50°C)
Power requirements
15 V @ 100mA
Options:
1.
Power
on
indicator
2. Remote detector test input multiplexer