Maintenance and Calibration
Maintenance
5
5-1
Section 5
Maintenance and Calibration
5.1 Maintenance
The UDR Simulator and signal converter are designed to operate for extended periods without scheduled 
maintenance. Yearly inspections of the UDR Simulator rear panel cable pigtails for chaffing is 
recommended. 
The 948B-1-5 Rack Chassis should be periodically inspected for dust build-up on the ventilation screens. 
Adjust the frequency of cleaning as required to maintain good convection airflow through the chassis. 
5.2 Calibration – Simulator Display
Calibration of the S14283A101 signal isolator is limited to the adjustment of the zero and span 
potentiometers. Refer to the Table in Section 4 for input vs. output voltages. 
Calibration of the display of UDR Simulator is limited to the adjustment of span on the 942-200-75 Analog 
Input Option circuit card. 
For additional calibration adjustments generic to the 956 series Ratemeter, refer to the 955A Standard 
Manual. Features covered include analog current loop output and voltage output zero and span. These 
features may or may not be required for simulator use. 