Change #3, Change #4 – Fluke Biomedical ESA609 Supplement User Manual
Page 3

Manual Supplement
ESA609 Users
2 12/14
To zero the resistance for units with the adapter, connect the adapter to the test receptacle. Follow the
procedure on page 13, under Performing a Ground-Wire (Protective-Earth) Resistance Test, and
replace the two bullets with:
•
If using an accessories probe, connect it to the other end of the test lead and place the probe tip into the
Ground Pin of the 15 A to 20 A adapter.
•
If using an alligator clip accessory, connect it to the other end of the test lead, place the null post adapter in
the Ground Pin of the 15 A-to-20 A adapter, and clamp the alligator clip to the null post adapter.
On page 29, under Specifications, replace the Power section with:
Power
115 Volt test outlet .............................. 100/120 V ac rms, 50Hz/60Hz, 20 A maximum
230 Volt test outlet .............................. 200/220/230/240 V ac rms, 50Hz/60Hz, 16 A maximum
Power input ......................................... 115V 20A - 2.6 kVA and 230V at 16A - 4.2 kVA
Change #3,
187
On page 30, under Leakage Current, replace the Accuracy with:
Accuracy
DC to 1 kHz ...............................
±
(1 % of reading + 1
μ
A)
1 kHz to 100 kHz ........................
±
(2.5 % of reading + 1
μ
A)
100 kHz to 1 MHz ......................
±
(5 % of reading + 1
μ
A)
Change #4,
188
On page 29, under Specifications, remove the Agency approvals.