Change #3, Change #4 – Fluke Biomedical DPM4 Supplement User Manual
Page 3

Manual Supplement
DPM4 Users
2
1/14
Change #3,
63745
On page 3, under Specifcations General change the Maxiumum Humidity, Operating:
Maxiumum Humidity, Operating ................... 80 % relative humidity up to 31 °C (88 °F), decreasing linearly to 50 % relative
humidity at 35 °C (95 °F)
Under External Power Supply change the Output Current:
Output Current ................................................. 2.5 A
Change #4,
66451
On page 2,in the Symbols table remove:
Manufacturer’s declaration of product compliance with applicable
EU directives
On page 5, under Temperature Measurement replace the following:
Units of measure ...................................... °C and °F
Use standard external temperature probe type PT-100 or PT-1000 (DIN/IEC 751 Class B) for temperature measurements in °C or °F.
PT-100 Temperature Probe
Operating range...................................... -50 °C to +200 °C
Accuracy ................................................. 0.3 °C @ 0 °C, 0.8 °C @ 100 °C
PT-1000 Temperature Probe
Operating range...................................... -50 °C to +200 °C
Accuracy .................................................
(0.3 °C +0.005 (t) °C)
Where t is the temperature, in Celsius, that is being measured.