Making measurements, Zeroing – Fluke Biomedical DPM2 Plus User Manual

Page 13

background image

DPM2 Plus Digital Pressure Meter

Making

Measurements

5

Table 1. Switch Positions

Switch Position

Function

Off

Turn off power to the Meter

mmHg

Pressure from -698 to +802 millimeters of mercury

cmH

2

O

Pressure from -949 to +1090 centimeters of water

inH

2

O

Pressure from -374 to +429 inches of water

-13.5~15 psi

Pressure from -13.5 to +15.50 pounds per square inch

-13.5 ~ 100 psi

Pressure from -13.5 to +100 pounds per square inch

Making Measurements

The Meter has a large 0.5–inch liquid crystal display (LCD) to indicate the
pressure at the Luer-Lok input. This male Luer-Lok provides the external
connection for stopcocks or pressure fittings.

Note

It may take a few seconds for the display to settle to zero when the
Meter
is initially turned on and vented to atmosphere.

The Meter has five different range measurement units as indicated on the front
panel overlay in Figure 1. Two of the ranges measure psi, one of which allows
measurements of up to 100 psi. This selection option allows the operator
greater accuracy at lower psi ranges while maintaining the option to read
higher psi levels.

Zeroing

Select the desired range for the pressure measurements to be taken.

For best accuracy, make sure the pressure meter reads zero when vented to
atmosphere. Allow the Meter display to settle (due to changes in environment
it may take a few seconds for the display to read zero).

If the display will not read zero, use a small screwdriver to adjust the
potentiometer located on the side of the Meter until the display reads zero.