Page 6

Nuclear Associates 07-492
Operators Manual
1-2
Correction factors for different target/filter
combinations are provided in Section 5, Calibration
and Adjustment.
The kVp meter uses two differentially filtered detectors whose ratio of integrated outputs is calibrated over
the range of 24 to 40 kVp. Measurement results are presented on an easy to read 3½ digit Liquid
Crystal Display. Status is indicated by decimal point positioning on the readout. Low battery condition is
indicated by LO BAT displayed on the readout.
1.2 Specifications
Range
Mo: 24 kVp to 40 kVp
Accuracy
± 1.5 kV
Reproducibility
± 0.5 kV
Resolution 0.1
kV
Distance Dependence Negligible for SSD of 20 cm or greater
Angle Dependence
Negligible for 10 degrees or less
Position Dependence Negligible for misalignment of 5 cm or less at 30 cm.
Calibration Period One
year
Filtration Effects
1.5 kV per mmAl at 25 kVp
4.0 kV per mmAl at 35 kVp
Minimum Requirements
75 mA @ 24 kVp, 25 cm
Power Requirements 9V Alkaline Battery, MN1604A or equivalent
Display
3 ½ Digit LCD
Low Battery Indication: LO BAT
High kV Indication: Left Blinking Decimal
Low kV Indication: Middle blinking Decimal
High Intensity Indication: Right blinking decimal and non-zero display.
Low mAs Indication: Right blinking decimal and zero display.
LED Indicator
Auto Reset, new exposure detected, and last reading cleared.
Controls
Front Panel: Power switch.
Rear Panel: Phase selection switch
Output Connector
BNC Connector on rear panel for waveform output.
Dimensions (H x W x D)
2.5 in. x 8 in. x 6 in. (6.4 cm x 20.3 cm x 15.2 cm)
Weight
2.1 lbs. (.95 kg)
Temperature Limits
Operating: +18° to + 40° C (+64° to + 104° F)
Maximum 90% relative humidity, non-condensing
Storage: -18° to + 55° C (0° to + 132° F)
NOTE