Introduction, 1 general description, 2 application – Fluke Biomedical 6000-528 User Manual
Page 5: 3 specifications

Introduction
General Description
1
1-1
Section 1
Introduction
1.1 General Description
The Model 6000-528 External Ion Chamber is a low profile ion chamber designed to measure diagnostic
x-rays from 30 to 150 kVp. The external ion chamber consists of an ionization chamber and a 15-foot
cable assembly.
1.2 Application
The Model 6000-528 External Ion Chamber is specifically designed to be used with the NERO® 6000B,
6000M, 4000+, or the 4000M+ for measuring exposure from diagnostic x-ray machines. It can also be
used with any properly configured charge measuring electrometer. The coaxial BNC on the chamber body
is the signal output. The recessed banana plug is the high voltage connection for chamber bias. The
middle hole is threaded with a standard 1/4 - 20 thread for accessory mounting.
1.3 Specifications
Chamber
Radiation Measured
X-rays from 30 to 150 kVp
Nominal Sensitivity
9 nC/R
Volume
Nominal 30 cc
Energy Response
30 to 150 kVp, within 7%
Wall
0.031 inch phenolic (0.8 mm)
Termination
Signal - coaxial, fully guarded BNC
High Voltage - recessed banana
Cable Length
15 ft (3.8 m)
Dimensions
4 in x 4 in x 0.54 in
(10.2 cm x 10.2 cm x 1.4 cm)
Weight
4 oz (113 grams)