General information, 1 specifications – Fluke Biomedical 35035 User Manual
Page 5

General Information
Specifications
1
1-1
Section 1
General Information
1.1 Specifications
CURRENT (mA)
MILLIAMP SECONDS (mAs)
Range Resolution
Input Impedance*
Range Resolution
Threshhold
Current
Input
Impedance*
20 mA
0.01 mA
100 Ohms
200 mA
0.1 mA
10 Ohms
200 mAs
0.1 mAs
3.2 mA
10 Ohms
2000 mA
1 mA
1 Ohm
2000 mAs
1 mAs
32 mA
1 Ohm
SIGNAL INPUT LIMITS
Function
Range Maximum Input Limit
Function
Range
Maximum Input Limit
mA
OFF Input shorted –2.0 A max
(fuse protected)
mAs
20
Input shorted –2.0 A max
(fuse protected)
20 250 mA for 30 sec*
200
1 A for 30 sec*
200 1 A for 30 sec*
2000
2.0 A max (fuse protected)
2000 2.0 A max (fuse protected)
* Limit set by power dissipation rating of shunt resistors. Maximum common mode voltage is 500 V.
GENERAL
Meter Accuracy: ±1% of rdg + 2d on all ranges Input
Jacks: touch - protected 4 mm banana
Battery Type: Eveready type 522 or equivalent
Fuse: 5 x 20 mm type F2AH250V fuse
Battery Life: Approximately 200 hours
Display: 3-1/2 digit LCD; low battery annunciator,
0.50 in high digits
Environmental:
Operating Temperature: 5°C to 35°C
Storage Temperature: -20°C to 50°C
Humidity: 0 to 80%
Test Leads: 91.4 cm (36 in.) safety set with finger
guards on probes and shrouded banana plugs
Dimensions and Weight:
15 cm wide x 5.0 cm deep
x 8.75 cm high (6 in x 2 in x 3-1/2 in);
0.35 kg (0.78 lb)
* Specifications subject to change without notice.