Fluke Biomedical 07-644 User Manual
Page 9

Operation
Operation
2
2-3
Sample Grid Alignment QC Form
Room
103
Unit
GE II
Grid
Tested
Radiographic Table
Grid Ratio
12:1
Line
No.
103 lines/in.
Focal
Distance
40”
SID
40”
kVp
60
mA 5 mAs Time N/A
Measured Optical Density
: : Reference Location
(3 small marking holes toward
the front of the table)
0
Date 2 1
Center
1 2
Comments
Checker’s
Initials
1-15-94 .61 1.11 1.47
1.14
.65 Alignment
OK
MR
5-18-94 .31 .73 1.17
1.47 1.05 Check
Alignment
MR
5-19-94 .63 1.13 1.51
1.15
.66
OK After Service
MR
See also other documents in the category Fluke Biomedical Equipment:
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-453 (24 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)