General information, 1 parts, 2 assembly – Fluke Biomedical 07-653 User Manual
Page 5: 3 measurement setup

General Information
Parts
1
1-1
CAUTION
Section 1
General Information
1.1 Parts
Phantom Base Assembly
Thin Clip-On Attenuator (X.X mm Cu)
Thick Clip-On Attenuator (Y.Y mm Cu)
1.2 Assembly
Pediatric patient simulation – phantom without attenuators
Small adults – phantom with thin clip-on attenuator
Larger adults – phantom with thick clip-on attenuator
Heavy adults – phantom with both clip-on attenuators
1.3 Measurement Setup
The image intensifier’s grid is either in-place or removed in accordance with the clinical protocol being
simulated.
The entrance surface of the image intensifier is
fragile. Additional attention is required if the grid is
removed.
Fasten the phantom to the entrance surface of the image intensifier. (In some cases, the clips on the
phantom will engage slots on the image intensifier assembly.)
Set systems with variable source to image receptor instance (SID) to a standard value (e.g. 100 cm).
Remove extraneous material from the x-ray beam.
The tabletop and /or patient mattress may be in place if desired.