Instructions, 1 instructions – Fluke Biomedical 07-601 User Manual
Page 7

Instructions
Instructions
2
2-1
Section 2
Instructions
2.1 Instructions
1. Place (or tape) the test tool in contact with the image amplifier base frame or film changer. For under-
table intensifier tubes, lay the test tool on the tabletop.
2. Operate in the fluoroscopic mode at the lowest kVp and mA possible. Collimate so only the pattern is
visible with the edge limited by the collimator.
3. The test may be repeated using a suitable phantom, such as the Nuclear Associates 07-706, as a
beam block to determine clinical resolution capability.
4. For cineradiographic systems, exposure of only a few frames is needed to evaluate system resolution.
Only persons qualified to work on image intensifier
x-ray systems should attempt to focus and align
them. Testing to the stated criteria will indicate
proper performance. If the system falls short of its
capabilities, contact qualified
service personnel.
CAUTION