Specifications – Fluke Biomedical 07-453 User Manual
Page 16

Section 6
Specifications
Model 07-453 SPECIFICATIONS
Accuracy
•
AC Input / X-ray Pulse - +/- 1 Count
•
DC Input - 2% +/- 1 Count
o
Measures at 70% of peak value
Sensitivity
•
AC Input - 90 VAC minimum
•
X-ray input - 50kVp, 5mA at 2.0 inches from top surface of case, pointed to
target on case.
Range
•
9999
Pulses
•
9999
Milliseconds
Display
•
0.3" Liquid Crystal
Power
•
9 Volt Battery
Low Battery Indicator
•
"Lo Batt" Appears in display when battery voltage reaches 5.3 Volts +/- 0.3 Volt.
Size
•
80 X 147 X 40 mm
•
3.15 X 5.8 X 1.6 inches
Battery Life
•
48 Hours minimum
o
With Duracell MN1604 Alkaline or equivalent
AC Input Jacks
•
130 Volts AC maximum
•
90 Volts AC minimum
•
Input circuit not affected by reversed polarity
Optional Accessories
•
Remote x-ray sensor
•
Carrying
case
•
AC test leads with tip plugs
- 7000DP Impulse Supplement (12 pages)
- 7000DP Impulse Getting Started (36 pages)
- 7000DP Impulse (92 pages)
- 06-526-2200 (26 pages)
- 07-417 (16 pages)
- 07-443 (14 pages)
- 07-451 (8 pages)
- 07-487 (12 pages)
- 07-555 (8 pages)
- 07-553 (10 pages)
- 07-605-7777 (8 pages)
- 07-444 (46 pages)
- 07-618 (10 pages)
- 07-600 (6 pages)
- 07-591 (8 pages)
- SigmaPace 1000 (154 pages)
- 07-653 (8 pages)
- 07-633 (20 pages)
- 07-649 (14 pages)
- 07-661-7662 (12 pages)
- 07-645 (14 pages)
- 10100AT (80 pages)
- 07-644 (10 pages)
- 18-203 (4 pages)
- 07-621 (12 pages)
- 07-647 (12 pages)
- 18-207 (8 pages)
- 18-216-1000 (10 pages)
- 18-220 (10 pages)
- 18-228 (8 pages)
- 18-223 (18 pages)
- 18-229-1313 (6 pages)
- 18-250 (8 pages)
- 18-252 (6 pages)
- 18-303 (6 pages)
- 35035 (19 pages)
- 6000-528 (18 pages)
- 35080B (70 pages)
- 6000-529 (16 pages)
- 35080M (64 pages)
- 6000-530B (16 pages)
- 57-436 (12 pages)
- 57-440 (12 pages)
- 57-402 (32 pages)