2 specifications – Fluke Biomedical 07-443 User Manual
Page 6

Nuclear Associates 07-443
Operators Manual
1-2
1.2 Specifications
Density Range:
0 - 4.00 Density
Accuracy:
+ 0.02 D over specified range
Reproducibility: + 0.01 D
Operating Conditions: 50° F to 104° F (10° C to 40° C)
Maximum 90% Relative Humidity (non-condensing)
Aperture:
2 mm (1 mm Accessory)
Measuring Length:
5.3 inches (135 mm)
Display:
3 Digit 1/2 inch LCD, w/Polarity and Low Battery Indicator
Zero Range:
Auto zeroes up to density 1.00 D
Sensor:
20
R/sec
High
Efficiency
Silicon
Photodiode
Controls:
Zero: Push-button; automatically zeroes unit
Power:
ON/OFF push-button turns unit power on/off
Calibration Control:
Screwdriver adjustment; 20 turn potentiometer used to calibrate against a know
step tablet.
READ Push-button: Initiates read sequence
LAMP Push-button: Turns light source on for three (3) seconds
Light Source:
Ultra bright lamp turns on during measurement; provides extremely long life with
minimal spectral and intensity degradation; reduces specimen heating to a
minimum.
Dimensions (H x W x D):
2.40 x 3.20 x 7.10 in. (61 x 81 x 180 mm)
Weight:
1.72 lbs. (78 g)
Power:
Four (4) 1.5 V AA Batteries
Supplied Accessory: Part No. 010128, 5 step precision wedge Part No. 166055, 1 mm Aperture
Certification:
Designed to measure diffuse transmission density per ANSI specification PH 2-
19-1986