Standard setting – Flowserve NRGS 11-1 User Manual
Page 14
14
The measuring range can be switch selected between
≥
0.5 µS/cm and
≥
10 µS/cm
by means of code switch :
1. Unscrew screws at the electrode terminal, remove cover , Fig. 6
DIP switch 1 – 4 OFF
Measuring range
≥
0.5 µ S/cm.
DIP switch 1 – 4 ON
Measuring range
≥
10 µ S/cm.
2. Replace cover and fasten screws .
Standard setting
Factory setting
The compact system is delivered with the following factory settings:
■
Measuring range
≥
10
µ
S/cm
Switch selection of measuring range
Attention
■
Screwdriver for cross head screws, size 1
■
Screwdriver for slotted screws, size 2.5; completely insulated according
to VDE 0680
■
Do not damage the electronic components when setting the code
switch.
Tools
5
E
G
E
G