beautypg.com

Troubleshooting apex 8000 positioner – Flowserve APEX 8000 High-Performance Positioner User Manual

Page 11

background image

Troubleshooting APEX 8000 Positioner

Failure

Probable Cause

Corrective Action

Valve won’t

stroke, no

excessive air is

exhausting from

positioner

1. Tubing to wrong ports

2. Cam action reversed

3. Feedback lever arm is stuck

4. Low air supply

5. Relay tube stuck

6. Balance adjust screw not adjusted correctly

7. I/P module filter plugged

8. I/P module failure

9. I/P mounting bolts loose

10.I/P pressure signal blocked

1. Re-tube to correct ports (see “installation section).

2. Refer to “installation” section and reverse cam

3. Work with lever arm until it turns freely

4. Increase air supply to recommended value

5. Disassemble relay assembly and work relay tube free. Lightly

lubricate if necessary

6. Adjust balance pressure with adjusting screw

7. Remove I/P module and replace filter

8. Replace I/P module

9. Tighten mounting bolts

10. Remove I/P module and clear passageway; replace O-ring if

necessary

Valve won’t

stroke, excessive

air exhausting

from positioner

1.

A diaphragm in relay assembly burst

2. One of the poppets is stuck

3. Internal control valve problem

4. Damaged relay O-rings on relay tube

5. Blocked passageways in relay

1. Replace relay assembly or replace diaphragms

2. Remove relay assembly/poppet cover; free stuck poppet

3. Refer to instructions or check for actuator tubing leaks

4. Disassemble relay and replace O-rings

5. Disassemble relay and check small holes under diaphragms; clean

if clogged

Actuator goes

to full signal

position

regardless of

signal

1.

Broken feedback spring

2. Linkage is disconnected or stuck

3. Orifice is clogged by water, oil or dust in air

supply

4. Bent flapper, damaged nozzle

5. I/P module failure

6. Clogged orifice screen

1. Replace feedback spring

2. Check and tighten bolts/nuts in linkage. Make sure linkage does not

hang up; grease ping that rides in follower arm slot

3. Remove orifice screw and carefully clean orifice hole

4. Straighten flapper or replace damaged parts

5. Replace I/P module

6. Remove relay and clean ore replace orifice screen

Calibration

1. Loose positioner mounting

2. Loose linkage

3. Loose zero adjustment locking knobs

4. Wear of arms or pins

5. I/P mounting bolts loose

6. Stroke has change in valve

1. Remove cover and check three screws holding positioner to bracket;

check two bolts holding bracket to yoke

2. Tighten nuts and bolts on linkage and stem clamp

3. Tighten zero adjustment locking knob; re-calibrate if necessary

4. Replace worm arms, pins; grease appropriately

5. Tighten I/P mounting bolts

6. Refer to valve maintenance instructions

Excessive air

consumption

(other than

normal exhaust)

1. Air leakage from manifold rings between relay

and base

2. Air leakage from tubing

3. Leaky cylinder piston O-rings

4. Air leakage from relay

1. Tighten screws holding relay assembly together and/or replace

O-rings

2. Tighten or replace tubing fittings

3. Replace O-rings in cylinder

4. Disassemble relay and check and replace dynamic O-rings next to

tube if necessary

Actuator strokes

very slowly in one

direction only

1. Connection between signal capsule and flapper

misadjusted

2. Tubing to cylinder is restricted

3. Balance pressure low

1. Adjust gain according to Figure 6 or until actuator strokes

approximately equal speed in both directions. Verify alignment of

upper and lower gain plates. Make sure spacer nut is tightened.

2. Inspect tubing/fittings for restriction and replace if necessary

3. Adjust balance pressure according to page 5

Erratic operation

1. Dirt buildup on relay poppets or seats

2. Dirt buildup on relay tube

3. Clogged ports / passageways in relay

4. Faulty I/P module

5. Clogged orifice screw

6. Mechanical binding in linkage or internal

galling in valve

7. Clogged orifice screen

1. Disassemble; clean poppets and seats; add air filter or change filter

2. Disassemble; clean relay and lightly lubricate; replace O-rings if

necessary; add air filter or change filter

3. Disassemble, inspect and clean all ports and passageways

4. Replace the I/P module

5. Remove orifice screw and carefully clean orifice

6. Tighten linkage or refer to valve maintenance instructions

7. Remove relay and clean or replace orifice screen

Excessive

overshoot

1. Restricted air flow to positioner

2. Balance pressure not set correctly

3. Gain is set too high

1. Adjust air supply as needed

2. Adjust balance pressure according to page 5

3. Lower gain mechanism until overshoot is minimized

Apex 8000 High-Performance Positioner AXENIM0130-06 11/10

11