beautypg.com

Bluelab EC Pen User Manual

Page 3

background image

3

4.0 Calibration

Calibration of conductivity is not required for this unit as it is factory calibrated.

However; if you wish to calibrate the unit follow the instructions below.

YOU MUST CLEAN THE PROBE BEFORE CALIBRATING.

See section 2.0.

Rinse probe in fresh water and place it in a known

standard solution. See chart below for the correct

solution.

Wait for reading to stabilize.

Hold down the cal button for 3 seconds until CAL

appears.

Release button and CAL P should be displayed. If Err is

displayed check the probe is clean and that the calibration

solution is fresh and uncontaminated.
A tick will appear on the screen to indicate that the calibration

was successful. The tick will disappear after 30 days. To reset

back to factory default remove/replace battery.

NOTE: If you need to test or calibrate in a 1500 ppm

solution, you MUST set the pen to EC, then multiply your

result by 540. If calibrating, multiply 2.8 by 540

(2.8 x 540 = 1512).
This unit DOES NOT measure in the 540 ppm scale.

1

2

3

4

EC

ppm 500

(TDS)

ppm 700

(EC x 700)

Solution value

2.77

1385

1940

Displayed value

2.8

1390

1940

1385

ppm

Hold for 3

seconds

5.0 Error messages

The following error messages appear for the following reasons.

Temperature

over range

Hardware

error

EC over range

Temperature

under range

Temperature

error

2.0 Cleaning and maintenance

Cleaning the EC pen probe periodically ensures accurate readings. The probe is

cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif” a trade name for a liquid

scourer cream used in home bathrooms and kitchens. Similar products are called

“Liquid Vim”, “Soft Scrub”, “Cif cream”, or “Viss”. Never use scented varieties as they

contain oils that contaminate the probe. Follow the steps below to clean the probe.

Remove shroud

Hold the body and pull the shroud off. Holding your

hand around the shroud for a few seconds will help

with removal.

Clean probe face

Place one or two drops of probe cleaner onto the probe

face and rub with your finger or Bluelab Chamois firmly

and vigorously, to clean the probe face.
If a heavy build up occurs around the temperature sensor

clean with a soft toothbrush to remove contamination.

Rinse probe

Rinse off all traces of cleaner under running water using

the same finger or other side of Bluelab Chamois.
Check that the water forms a film on the probe face with

no “beads” of water. If beading is present repeat the

cleaning process.

Replace shroud and test in a known solution to

ensure the unit has been adequately cleaned.

3.0 Battery replacement

The EC pen is powered with 1 x AAA alkaline battery. Do not use rechargeable

batteries. A low battery warning is indicated by a battery symbol appearing on the

screen. Only remove the battery cap when the batteries require changing. Battery

life is expected to be 350 hours. Follow these steps to replace the battery.

Remove old battery

Undo battery cap fasteners. Remove battery cap and tip out

the old battery.

Check for corrosion

Flat batteries may leak and cause corrosion. Check battery

contacts and the battery for any sign of corrosion. Battery

contacts should be cleaned first if corrosion is found before

proceeding to step 3.

Fit new battery

Insert the new batteries positive (+) end down into the body.

Ensure waterproof battery cap seal is clean

Waterproof seal will fail if any dirt is around the seal and

body where the cap seals.

Replace battery cap

Tighten fasteners on battery cap until there is no space

left between the cap and body. This ensures the unit

remains 100% waterproof.

1

2

3

4

1

2

3

4

5

Pull

Waterproof seal