OPERATING MODES

Mode Selection:

By front panel rotary switch.

Mode A:

Independent outputs each capable of 0-75V at 0-2A.

Mode B:

Output 2 only active, 0-75V at 0-4A.
Output 1 disabled.

Mode C:

Output 2 only active, 0-150V at 0-2A.
Output 1 disabled.

METER SPECIFICATIONS

Meter Types:

Dual 3 digit meters with 14mm (0.56") LEDs. Reading rate 4 Hz.

Meter Resolutions:

100mV, 10mA

Meter Accuracies:

Voltage 0.3% of reading ± 1 digit,
Current 0.6% of reading ± 1 digit

Operating Modes:

Output 1 meters are blanked in Modes B and C.

GENERAL

AC Input:

110V-240V AC ±10%. Installation Category II.

Power Consumption:

500VA max.

Operating Range:

+5ºC to +40ºC, 20% to 80% RH.

Storage Range:

−40ºC to + 70ºC.

Environmental:

Indoor use at altitudes up to 2000m, Pollution Degree 2.

Safety:

Complies with EN61010-1.

EMC:

Complies with EN61326.

Size:

260 x 160 x 320mm (WxHxD).

Weight: 4.3kg