Specifications
Range:
1.0 x 10
3
to 1 x 10
12
@ 10V
1.0 x 10
7
to 1 x 10
12
@ 100V
Accuracy:
1.0 x 10
3
to 9.9 x 10
6
@ 10V 10% of reading at 70-80
0
F; 15–60% RH
1.0 x 10
7
to 1 x 10
12
@ 100V 10% of reading at 70-80
0
F; 15–60% RH
Open Circuit Voltages:
10 and 100 volts
3%
Electrification Period:
15 seconds
Temperature Accuracy:
5 degrees F and 3 degrees C Typ.
Relative Humidity:
Range from 5% - 95%;
10% Typ.
Power:
Two - Alkaline AA batteries
Battery Life: approximately 1500 measurements.
Meter Jacks:
Left jack on meter accepts a 3.5 mm plug
Right jack on meter accepts a standard banana plug
Test Leads:
Black lead terminated with a 3.5 mm plug one end and a standard banana plug on
the other
White lead terminated with a standard banana plus both ends
External Electrodes:
Two NFPA-99 5-pound electrodes ( 80 oz.)
2oz with Shore A (IRHD) 50 – 70
durometer conductive pads that comply with EOS/ESD S4.1
Resistivity Electrodes:
Parallel conductive silicone rubber electrodes 4 inches long and spaced 4
inches apart mounted onto an adapter sled
Meter Weight:
15.5 oz. With out adapter; 23 oz. with adapter
Dimensions:
8” L x 4.3” W x 1.6” H
Kit includes:
Meter
Protective carrying case
Test leads
Two NFPA-99 5-pound electrode weights
Two AA Alkaline batteries
Operator’s Manual
Resistivity adapter sled is optional and must be ordered separately.