Monroe Electronics Electrostatic Fieldmeter - Hand Held - Intrinsically Safe - model 282IS User Manual
Page 4
Maintenance
Battery Care The battery should be replaced annually, whenever you plan an extended period
of unattended monitoring, or whenever “BAT” appears in the display for more than an instant.
See Specifications for replacement type. If unit fails to operate after replacing the battery,
remove battery, press and release the ON/ZERO button then re-connect the battery. Remove
the battery before long-term storage.
Cleaning To obtain accurate, drift-free readings the sensor plate and especially the area around
the aperture must be kept absolutely clean at all times. Always keep the instrument in its
protective case when not actually taking measurements. Never touch the aperture with
anything-not even cotton swabs. To remove dust or other particulate matter, use low-
pressure instrument grade air. To remove severe contamination spray or flush with the smallest
practical amounts of clean technical grade isopropyl alcohol. The stand the instrument on the
aperture end and allow it to dry for several hours. DO NOT clean with any solvent on a regular
basis.
Zero Drift Apply anti-stat solution to plastic surrounding the sensor plate using lint free
applicator. DO NOT allow entry into aperture.
Safety Considerations
The Model 282 I S has undergone testing to qualify for the following CE and UL markings:
CE — Radiated and ESD immunity to severity levels for industrial devices according to EN
50082-1 and radiated emissions with limits for Group 1, Class A product of EN 55011 to the
generic standard EN 50081-2 (1992).
UL — Certification for use in Class I, Division 1, Groups D, C, B, and A and Class II, Division 1,
Groups G, F and E hazardous areas when powered by a 9-volt carbon/zinc or zinc chloride
battery, NEDA 1604 or NEDA 1604D and for use in Class I, Division 1, Groups D and C
hazardous areas only, when powered by a 9-volt alkaline battery.
Two-Year Warranty
Monroe Electronics, Inc. warrants this instrument to be free from defects in material and
workmanship for a period of two years after shipment from the factory. This warranty is
applicable to the original purchaser only. Liability is limited to service, adjustment or
replacement of defective parts (other than fuses or batteries) on any instrument returned to the
factory, transportation charges prepaid. It does not apply to any instrument subjected to abuse,
abnormal operating conditions, or unauthorized repair or modification. No other warranties,
explicit or implied, apply. Under no circumstances shall Monroe Electronics be responsible for
damages either direct or consequential. For additional information concerning this product,
contact Monroe Electronics, Inc.
Copyright
1999 Monroe Electronics, Inc.
Specifications subject to change without notice.
Stat-Arc is a trademark of Monroe Electronics, Inc.
Printed in USA