Development kit for DATAMAN 570 series

Programmer’s Guide

- 22 -


Return value: -

3.5. Other functions

3.5.1. GroundPositionToShift

Returns vertical shift, which is necessary to set, to have the 0V at the selected position
of the screen (quantization level).

int __stdcall __export GroundPositionToShift(int channel, int position);


Parameters:

channel – determines channel

CHANNEL_A – channel A

CHANNEL_B – channel B

position – screen position (quantization level), where the 0V should be placed


Return value:
The vertical shift value, which places 0V to the selected screen position (quantization
level). Set this value using SetVert to place 0V to the selected position.