beautypg.com

American Weigh Scales (AWS) MSC-500 User Manual

Page 4

background image

obtain just the weight of the contents.

5. When the container is removed, a negative number
will be displayed. Press [ ] again to zero the scale.

IV. CALIBRATION

This scale has come pre-calibrated at the factory. Only

re-calibrate if you are experiencing trouble weighing.

MSC-100 uses a 100g calibration weight

MSC-500 uses a 500g calibration weight

A. MSC-100 Calibration:

1. Turn the scale OFF.
2. While holding the [ M ] key, tap the [ ] key to turn

the scale on then release both keys.

3. The display will show a random number, press the

[ M ] key again.

4. The display will flash “100.00”. Place the required

weight on the platform and wait for 3 seconds.

5. The display will show “PASS” and “100.00”. Remove
the weight and press [ ] to turn the scale off.

Calibration is complete!

B, MSC-500 Calibration:

1. Turn the scale OFF.
2. While holding the [ M ] key, tap the [ ] key to turn

the scale on then release both keys.

3. The display will show a random number, press the
[ ] key again.

4. The display will show “Zero” followed “500.0”. Place

the required weight on the platform then press the
[ ] key again.

5. The display will show “CAL PASS”. Remove the