beautypg.com

American Weigh Scales (AWS) DIA-20 User Manual

Page 6

background image

DIA-20

then release.

3. 3ress the 02'( keyDJDLQ

The display will flash CAL followed by “10.000g”.

4. Now place a 10g calibration weight on the platform

and wait 3 seconds, the display will flash "20.000g.

5. At this point, place another 10g on the platform to

make a total of 20g on the scale and wait a few

seconds.
6. The display will read “PASS” and then return to normal

weighing mode, calibration is finished! You can turn off
the scale.

X. Troubleshooting

Problem: Scale will not turn on

Solution: Check the batteries

Problem: Displayed weight fluctuates randomly

Solution: Make sure your workspace is stable and

free of any air currents or vibration

Problem: Displayed weight is inaccurate

Solution: Perform a calibration adjustment

XI. Error Codes

1. OUTZ - Zero range has shifted. Perform calibration

adjustment. If the problem persists, the load cell may

be permanently damaged due to overloading.

2. 88888 - Maximum capacity exceeded. Remove the