BS-Series Manual

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

BS-100

0.02g

100g

OIML M2

BS-250

0.2g

200g

OIML M2

IX. Adjustment

1. Press and hold the

M key until the display shows

“CAL”.

2. Press the

M key again. The display will flash “CAL”

followed by the calibration weight value (see chart).

3. Place the calibration weight on the center of the

weighing platform. Make sure you are using the

correct weight for your specific model.

4. Once the weight is accepted, the display will show

“PASS” then return to normal operation.

X. Troubleshooting

Problem: Scale will not turn on

Solution: Check the batteries

Problem: Displayed weight fluctuates randomly

Solution: Make sure your workspace is stable and

free of any air currents or vibration

Problem: Displayed weight is inaccurate

Solution: Perform a calibration adjustment

XI. Error Codes

1. OUTZ - Zero range has shifted. Perform calibration

adjustment. If the problem persists, the load cell may