remove any items from the weighing platform and

press the

ON/T key again to re-zero the scale.

VII. Calibration
Calibration and adjustment are performed at the factory.

However, the weighing range can shift slightly during

shipping. Also, due to changes in local acceleration of gravity

around the world, adjustment is recommended when moving

the scale from one location to another. To check calibration:

1. Place the required calibration weight on the scale

and note the displayed value.

2. If the displayed value is not within acceptable

tolerance (see chart below), adjustment is required.

Model

Acceptable

Tolerance (±)

Calibration

Weight

Recommended

Accuracy Class

CM-500

0.2g

500g

OIML M2

VIII. Calibration Adjustment

1. With the scale off, press and hold the

MODE and

ON/T keys at the same time until the display shows

a random number then release.

2. Press the

ON/T key again and the display will flash

“AL_F” followed by “500.0g”.

3. Place a 500g calibration weight (not included) onto

the weighing platform,

4. Once the weight is stable, press the

ON/T key once

more and the display will show “AL_F” and return to

normal weighing mode. Calibration is complete.