Tcat-2 controller, 0 operating instructions, Operating instructions – Physitemp Instruments TCAT-2AC User Manual

Page 7

background image

4.0

OPERATING INSTRUCTIONS

The Instruction Label from the top of the instrument is reprinted here in case it is ever defaced.

4

OFF/ON
POWER

SWITCH

FEEDBACK

SENSOR

CONTROLLED

AC OUTPUT

TCAT-2

CONTROLLER

OPERATING INSTRUCTIONS

The TCAT-2 Controller parameters are set at the factory for the specific load supplied with it. If the
TCAT-2 has not been so supplied, see the Operating Manual for adjustment of the control parameters
1.

Connect the thermocouple sensor to the blue Feedback Sensor Socket on the rear panel and

position the sensor.

2.

Connect the load (lamp or mat, etc.) to the Controlled AC Output on the rear of the controller

(7amp max.) or the Low Voltage 7-pin Output Socket

3

Connect the controller power cord to a suitable AC outlet.

4.

Turn on controller Power Switch (rear panel). Set position of Heat/Cool Switch depending on

control protocol (See manual for details.)

5.

Set the control temperature:

a) Press SHIFT. A digit will flash.

b) Use UP and DOWN buttons to change

the flashing digit. Press SHIFT again to

select the next digit.

c) When all digits are changed, press SET

to enter new control temperature.

SPECIFICATIONS:
Operating Range:

-200°C to +400°C

Resolution:

0.1°C

AC Input Voltage:

110-130VAC or 220-240VAC, 50 or 60Hz (see Operating Manual)

SPECIAL DISPLAY INDICATIONS:
Red lamp will be illuminated when power is being supplied to load.

VIEWING ANGLE
The two front feet can be flipped down to adjust the viewing angle. Rotate feet down and

towards the front until they lock in position. To return to original position, pull down slightly,

then rotate back.

Made in USA

SET

UP

DOWN

Serial Number:

SV -15.0

PV -14.9

SHIFT

PHYSITEMP INSTRUMENTS INC

154 Huron Avenue

Clifton, NJ 07013 USA

Tel: 973-779-5577, Fax: 973-779-5954

RUN Display

SET Display

9003