4.0 SPECIFICATIONS
4.1 CONTROLLER
Operating Range:
-20
o
to
+
60
o
C
Control Accuracy:
+
0.1
o
C
Digital Readout Resolution:
0.1
o
C
Accuracy: 0.1
o
C,
+
1 digit
Ambient Operating Range:
15
o
- 45
o
C
Input Power Requirements:
100 - 120VAC, 60Hz, 100W
200 - 240VAC, 50Hz, 100W
Safety Features:
Safety shutdown with warning lamp in case of fault condition
such as sensor breakage, lack of cooling water or electronics
failure.
Size:
8” high x 7” wide x 15” deep
Weight
8 lbs.
Other Features:
Spring loaded switch indicates set point
Self-sealing water connectors
Auxiliary AC switched output
4.2 STAGE
Temperature Range:
-20
o
to +60
o
C
Controlled Surface Area:
1 1/4” x 1 1/2”
Weight:
16 ozs., including leads
Material:
Copper, nickel plated
Mounting:
Plastic. Fits slide holder
4.3 PUMP AND
TANK
UNIT
Size:
6 1/2” x 15” x 14”
Power Supply:
110V (or 220VAC if ordered)
Power Consumption:
150 Watts
Water Connections:
Self Sealing
4.4
THERMOMETER (built into controller)
Temperature Range:
-100
o
to +200
o
C
Resolution:
0.1
o
C
Sensor:
Type T Thermocouple, Physitemp type #MT-26/4 or similar
7