LW Measurements WB 10000 User Manual
Page 13
. Output: 12V DC 150mA
!
Allowable ambient conditions
!
Temperature: 5°C - 40°C
Relative humidity: 25% - 85%, non-condensing
!
If you have any questions on the technical data or require detailed technical information on your balance,
please contact your technical representative.
!!
2.3.2 RS232 Data interface
!
1.
Connector: DB9
!
SCALE
PC
2
2
3
3
5
5
!!
2.
Transmissions Settings
!
Mode: Simplex Asynchronous Serial
Baud Rate: 9600
Data Bit: 8
Parity Bit: None
Stop Bit: 1
Data Format: ASCII
!
3.
Transmission Information Format: 20 Byte ,blank=20H
! !
W:
Start of Data Transmission=57H+3AH
P :
Polarity ‘+’ = 2BH = Positive
‘-’ = 2DH =Negative
DATA: ‘ 12.345’= 20H+20H+20H+20H+31H+32H+2EH+33H+34H+35H
UNIT: ‘lb’
=6CH+62H+20H+20H+20H
‘kg’
=6BH+67H+20H+20H+20H
CR:
= 0DH
LF:
= 0AH
!
1~2
3
4~13
14~18
19
20
‘W: ’
P
DATA
UNIT
CR
LF
!
13