LW Measurements HRB 1002 - 1000g x 0.01g User Manual
Page 20

20
5 Calibration - Using an External Calibration Weight
Calibration is required when the weighing machine is initially installed or if the Balance is moved
to a substantial distance from the original location. 30 minutes of warming up of the Balance is
always needed before weighing and calibration.
Turn the Balance on and when display shows
“--------”, press ZERO key
The display will show “CAL”
Press ZERO key again, the display will show “X0000 CAL”, “X” is the flashing digit
Press UNIT key, the flashing digit will move to right
Press TARE key t
o increase the value of the flashing digit(“X” is the calibration weight
can be set according to users’ requirements, we recommend a minimum weight of at
least 50% of the Balance capacity)
Place the corresponding test weight on the center of the pan
Press ZERO key to confirm and the display shows the weight value
Now the calibration is completed
Turn off the power, then turn on the power again, place a known weight on the platform
to verify if it is accurate.
If not, repeat above steps.