2 rs232 data interface – LW Measurements LVS 700 XL User Manual
Page 12
![background image](/manuals/739198/12/background.png)
12
2.3.2 RS232 Data interface
1. Connector: DB9
SCALE
PC
2
2
3
3
5
5
2. Transmissions Settings
Mode: Simplex Asynchronous Serial
Baud Rate: 9600
Data Bit: 8
Parity Bit: None
Stop Bit: 1
Data Format: ASCII
3. Transmission Information Format: 20 Byte ,blank=20H
1~2
3
4~13
14~18
19
20
‘W: ’
P
DATA
UNIT
CR
LF
W:
Start of Data Transmission=57H+3AH
P :
Polarity
‘+’ = 2BH = Positive
‘-’ = 2DH =Negative
DATA:
‘ 12.345’= 20H+20H+20H+20H+31H+32H+2EH+33H+34H+35H
UNIT:
‘lb’
=6CH+62H+20H+20H+20H
‘kg’
=6BH+67H+20H+20H+20H
CR:
= 0DH
LF:
= 0AH