LW Measurements LDS Large Drum Scale User Manual
Page 17

17
3.4 Choice of a suitable location
The environment in which your weighing machine is used is 
very important. Air movement, temperature changes, 
vibrations, direct sunlight, etc. all influence the performance of 
high precision weighing machines. Therefore, place your 
weighing machine on a solid, sturdy surface that is free of air 
currents, vibration and not in direct sunlight. The surface 
should not be magnetic and should be located away from 
doors, windows, heaters, air conditioners and fans. 
 
To summarize: 
 
Put the weighing machine on a solid, firm and 
preferably vibration-proof, horizontal base 
Make sure that the weighing machine cannot be 
shaken or knocked over 
Protect from direct solar radiation
Avoid drafts and excessive temperature fluctuations
3.5 Checking the mains voltage
The following Safety recommendations must be observed 
when connecting the balance to the mains: 
 
 
The balance may only be operated with the power adapter 
supplied. 
 
Check before connecting the power adapter to the mains 
supply, that the operating voltage stated on the power adapter 
agrees with the local mains voltage. 
 
