FUNCTION SETTINGS
1. Press and hold [HOLD] key then press [ON/OFF] key to turn on the scale, the
display will show A-ON or A-OFF, press [ZERO] key to select on or off (ON
means auto power off function activate, OFF means auto power off function
inactivate).
2. Press [UNIT] key, the display will show L-ON or L-OFF, press [ZERO] key to
select on or off (ON means auto backlight function activate, OFF means auto
backlight function inactivate).
3. Press [UNIT] key, the display will show P-XX. Press ZERO key to select OFF,
key, con and Stb. (OFF means RS232 function inactivates, KEY means that
output the data by pressing [PRINT] key. Con means that output the data
continually. Stb means that output the data when reading is stable)
4. Press [UNIT] key, the display will show J-ON or J-OFF, press ZERO key to
select ON or OFF. (ON means showing zero when turn on the power, OFF
means showing weight placed when turn on the power.)
5. Press [UNIT] key, the display will show F-Nor, F-PEA or F-Ani. Press ZERO
key to select Nor or PEA or Ani. (Nor means normal weighting function activate.
PEA means peak hold weighting function activate. Ani means animal weighting
function activate).
6. Press [UNIT] key, the display will show N-dsp or N-Ind. Press ZERO key to
select dsp or Ind. (dsp means it can display synchronously as a displayer for
another indicator through the RS232 socket. Ind means it is only a indicator for
the scale)
7. Press [UNIT] key, the display will show Kd 50 or Kd 100 or Kd 200 or Kd400.
(press [ZERO] key to select animal scale anti-shock range among 50, 100, 200,
400)
3
Features
:
1. Display options:5 digital LCD
2. Power:AC adapter 9V 100mA
3. Auto back light (selectable)
4. Auto off (selectable)
5. Operating Temperature:0-40
℃(32-104℉)
6. Operating Humidity:25%~95%RH
7. 5 Keys:ON/OFF, ZERO, UNIT, HOLD, PRINT, and special foot switch for
TARE function
ZERO
This key is “Zero” feature if load is below 10% of full
capacity or a negative value
UNIT Toggle among different units
ON/OFF Turns the scale power On or Off
HOLD Press this key to store weight value
PRINT Press once and send weight information once.
8. Zero Range (0 to 10% of full capacity)
9. Tare Range (10% to full capacity)
10. Error Message indication:
Err-0 Overload
Err-5 Unstable when power on
Err-c Calibration code is not enough or operation error.
Low battery indication
11. Two calibration weight units (Optional): kg , lb
12. Open calibration capacity
13. Zero range: + 10% of full capacity
14. Three modes : Normal mode / Setting mode /Internal Auto Calibration
mode
15.
RS232 function
8