Laerdal Multi-Venous IV Training Arms User Manual
Page 2

Table of Contents
Introduction .............................................................................................................2
Laerdal Recommends ............................................................................................2
Items Included .........................................................................................................2
Skills Taught ..............................................................................................................2
Anatomy Represented ...........................................................................................2
Instructions for Use ...............................................................................................2
Replacing Skin ..........................................................................................................2
Replacing Veins ........................................................................................................2
Care and Maintenance ...........................................................................................3
Replacement Parts .................................................................................................3
Introduction
The Multi-Venous Access Arm Kit (270-00001) is a full-size, lifelike
arm
reproduction of a male patient. It is specifically designed for training
professionals in the practice of venipuncture.
This trainer should provide many sessions of training with reasonable
care and maintenance after use.
Laerdal Recommends
Needle or Catheter - Size 21-gauge or smaller
For easier insertion, lubricate catheters or other venipuncture devices
with Manikin Lubricant spray
Caution: Latex
This product contains Natural Rubber latex which may cause
allergic reactions when in contact with humans.
Items Included
(1) Multi-Venous Access Arm
(5) Syringes
(1) IV
Bag and Tubing Kit
(1) Replacement
Arm Skin
(1) Replacement
Multi-vein System
(1)
Red Simulated Blood Bottle
(1) Manikin Lubricant
(1) Carry Case
(1) Laerdal Global Warranty Booklet
(1) Directions for Use
Skills Taught
• Adult Venipuncture
Anatomy Represented
• Dorsal
Veins of Hand (3)
• Cephalic
• Median
• Basilic
• Antecubital
Instructions for Use
1. Attach IV tubing to either vein.
2. Allow fl uid to fl ow through arm and out other end of latex vein.
3. Clamp off fl ow of fl uid.
4. Tie a knot in the tubing exiting the arm (not attached to the bag).
Note: Knot should be tied so it will be easy to undo.
5. Open the fl ow of fl uid.
Trainer is now ready to practice venipuncture.
Laerdal Recommends
We recommend nothing larger than a 21-gauge needle. This will
extend life of skin and veins.
When using a catheter or other venipuncture devices, lubricate them
with lubricant spray for easier insertion.
Replacing Skin and Veins
Laerdal Recommends
When excessive leaking occurs at the puncture sites, a new vein system
should be installed to reduce loss of fl uid.
We recommend working at a basin when replacing skin and veins.
Replacing skin:
1. Cut off skin. This can be done with a sharp knife or scalpel. (Fi
g1)
2. Discard skin.
Fig. 1
Replacing veins (
keeping skin):
1. Lubricate inside skin with liquid detergent; let it fl ow down into the
fi nger area.
2. Begin at top of arm and slowly pull skin down and off of arm. Do not
roll, as that will cause skin to bind. Thumb will detach with skin.
3. Remove tubing from track in mandrel. Glue may need to be scraped away
to allow removal.
4. Rinse and dry vein grooves well and swab with alcohol. Be sure to
remove any excess glue.
Fig. 2
5. Place new veins along grooves
(Fig. 2), spot gluing as needed. (We
recommend a fast-drying glue)
6. Generously lather arm mandrel with liquid soap.
E
NGLISH
Laerdal
2
Male Multi-Venous Access Arm Kit