Laerdal HeartSim 200 User Manual
Page 13

ENGLISH
5.1 Trouble Shooting
The Heartsim 200 is designed to be a reliable unit requiring a
minimum of maintenance. With this in mind, there are only a few
faults the user can rectify. In all other cases contact Laerdal
Medical, or your Laerdal dealer for authorized servicing or repair.
Problem
Correction
Battery Low LED is on
- Replace batteries
No ECG signals displayed
- Check all connections
No pulse simulation
- Check to see if battery
low LED is on.
- Check manikin cable
connection.
- Check type of manikin
(see page 4).
- ALS Skilltrainer only:
Pulse is pressed down on
both sides simultaneously.
5.2 Servicing
5.2.1 Battery Removal/Replacement
To remove and replace the batteries:
- Remove the battery cover on the rear of the Heartsim 200.
- Remove and safely discard the old batteries.
- Install the new batteries, AA size (R6) 1.5V, in the correct
orientation as shown on the cabinet bottom.
- Re-fit the battery cover.
5.3 Cleaning
5.3.1 General
Cleaning of the Heartsim 200 is restricted to general cleaning of
the keyboard and case:
- Moisten a cloth with a suitable cleaning agent. Use a mild soap
or cleaner.
- Gently wipe the surface of the keyboard and case.
- Remove any excess liquid with a dry cloth.
- Do not immerse the unit in water.