Service operations – Labconco Protecto Series Evidence Drying Cabinets 3400004 User Manual
Page 32
Chapter 6: Maintaining the Cabinet
Product Service 1-800-522-7658
28
Service Operations
Door Evidence Prefilter Replacement:
1. Pull the prefilter frame(s) straight away from the door(s), as shown in
Figure 6-1.
2. Wearing appropriate gloves, and working on the inside surface of the
prefilter frame(s), carefully pull out the Evidence prefilter(s). The
Evidence prefilter(s) can now be examined for trace evidence, or stored
with the evidence in a sealed bag for future reference.
3. Examine the door prefilter(s). If it is contaminated from the evidence
drying operation, or from dust, replace it with a new one, such that the
orange side of the prefilter(s) faces you (away from the front of the
prefilter frame).
4. Press the prefilter frame(s) back into the door, engaging the fastener
strips.
Door Prefilter Replacement:
1. Follow Steps 1, 3, and 4 above.
Figure 6-1