HEPA Filter Replacement:
1. Remove the prefilter as described previously.
2. Using a nut driver or wrench, remove the HEPA filter clamps, as
shown in Figure 6-2.
3. Pull the HEPA filter away from the filter frame, and install the
replacement filter, with its gasket facing the filter frame.
NOTE: Take care not to touch the HEPA filter media. It is very fragile.
4. Reinstall the HEPA filter clamps, the prefilter, and the filter cover.