Chapter 5: Using The Ductless Enclosure
Product Service 1-800-522-7658
37
Work Techniques
Keep all materials a minimum of 4 inches (100 mm) from the inside
of the lower air foil, and perform all contaminated operations as far to
the rear of the work area as possible.
Segregate all clean and contaminated materials in the work area.
Arrange materials to minimize the movement of contaminated
materials into clean areas.
Keep all discarded contaminated material to the rear of the work area.
Avoid moving materials or the operator's hands and arms through the
front access opening during use.
Avoid the use of an open flame.
Avoid using techniques or procedures that disrupt the airflow patterns
of the enclosure.
Final Purging
Upon completion of work, the enclosure should be allowed to operate
for two to three minutes undisturbed, to purge airborne contaminants
from the work area.
Unloading Materials and Equipment
Objects in contact with contaminated material should be surface
decontaminated before removal from the enclosure.
All open trays or containers should be covered before being removed
from the enclosure.
Wipe-Down
Wipe down the interior surfaces of the enclosure with suitable
solvent, disinfectant, or cleaner. Important Note: Clean up
procedures using alcohols or volatile chemicals with low filter
capacity could saturate the filters quickly.
Dispose of used rubber gloves appropriately, and have
soiled/contaminated lab coats laundered properly.
Wash hands and arms thoroughly with soap.
Shutdown
Turn off the fluorescent light and enclosure blower to preserve filter life.