Electrical requirements, Space requirements – Labconco Class I Enclosure Models 3980423 User Manual
Page 11

Chapter 2: Prerequisites
Product Service 1-800-522-7658
7
Enclosure
Width
Model Description
Face
Velocity
(fpm)
Exhaust
Volume
(CFM)
Noise
Pressure
db(A)
Max.
Equivalent
Resistance
of 6" Duct
(Ft)
Max.
External
Static
Pressure
2'
2' Purifier Class I Enclosure
2' Purifier HEPA Filtered
Enclosure
60
75
80
90
100
105
85
110
115
130
145
155
48-53
49-55
50-56
51-57
53-58
54-59
400
220
200
140
80
70
.20"
.19"
.18"
.16"
.13"
.12"
3'
3' Purifier Class I Enclosure
3' Purifier HEPA Filtered
Enclosure
60
75
80
90
100
105
130
165
175
200
220
230
48-53
52-56
53-57
55-58
58-61
60-64
130
65
50
35
25
20
.15"
.12"
.11"
.09"
.08"
.07"
4'
4' Purifier Class I Enclosure
4' Purifier HEPA Filtered
Enclosure
60
75
80
90
100
105
175
220
235
265
295
310
49-54
54-58
58-61
60-63
62-66
64-67
100
60
50
35
25
20
.20"
.19"
.18"
.16"
.13"
.12"
Proper blower selection can be determined from these exhaust requirements and
the total system static pressure loss. For outside exhaust, the enclosure must be
connected to either a dedicated blower or a house exhaust system.
Labconco offers accessory remote blowers listed in Chapter 7. Contact Labconco
for blower sizing assistance.
If the enclosure is connected directly to a house exhaust system,
an adjustable damper (or valve) must be installed to control the
airflow properly. This is equally important when a house
exhaust system is controlling multiple filtered enclosures. See
Chapter 7 for accessory adjustable damper ordering
information.
Electrical Requirements
Standard duplex electrical receptacles should be nearby for connecting the filtered
enclosure, or other equipment. The enclosures include iris pass-throughs to allow
electrical cords through the back of the enclosure without leaving a large hole for
contaminants to escape.
Space Requirements
The dimensions for the different models are shown in Appendix B: Dimensions.
!
- Class I Enclosure Models 3980421 Class I Enclosure Models 3980403 Class I Enclosure Models 3980401 Class I Enclosure Models 3980323 Class I Enclosure Models 3980321 Class I Enclosure Models 3980303 Class I Enclosure Models 3980301 Class I Enclosure Models 3980223 Class I Enclosure Models 3980221 Class I Enclosure Models 3980203 Class I Enclosure Models 3980201 HEPA Filtered Enclosure Models 3980422 HEPA Filtered Enclosure Models 3980420 HEPA Filtered Enclosure Models 3980402 HEPA Filtered Enclosure Models 3980400 HEPA Filtered Enclosure Models 3980322 HEPA Filtered Enclosure Models 3980320 HEPA Filtered Enclosure Models 3980302 HEPA Filtered Enclosure Models 3980300 HEPA Filtered Enclosure Models 3980222 HEPA Filtered Enclosure Models 3980220 HEPA Filtered Enclosure Models 3980202 HEPA Filtered Enclosure Models 3980200