Now that the installation is properly prepared, you are ready to inspect,
install, and certify the Purifier Biosafety Cabinet. This chapter covers how to:
• Unpack and move the Biosafety Cabinet.
• Install the cabinet.
• Connect the electrical supply source.
• Connect the service lines.
• Connect to an exhaust system.
• Arrange certification of the Biosafety Cabinet.
Tools required for installation the Biosafety Cabinet include two 1/2"
wrenches, a flat-blade screwdriver, a #2 Phillips screwdriver, and a
carpenter’s level.
Note: The Biosafety Cabinet models weigh between 400–700 lbs (182-318 kg).
The shipping pallet allows for lifting with a mechanical lift truck or floor jack. If
you must lift the product manually, use at least six (6) persons and follow safe-
lifting guidelines.
Note: The side panels must be removed to access the fasteners that secure the
Biosafety Cabinet to the pallet. DO NOT ATTEMPT TO LIFT THE
BIOSAFETY CABINET BY THE SIDE PANELS; DAMAGE WILL
OCCUR.