Konig Electronic Blood pressure monitor wrist User Manual
Page 5

5
9. TROUBLESHOOTING
PROBLEM
POSSIBLE CAUSE
SOLUTION
LCD Display
shows battery
symbol
Low Battery
Change all batteries
LCD Display
shows “EE”
Arm or blood pressure monitor was moved
during testing
Re-test taking care to not move your arm or the blood
pressure monitor
The cuff does not inflate properly or pressure
falls quickly during testing
Ensure that the rubber tube is fully inserted into the
blood pressure monitor
Irregular heartbeat (arrhythmia)
People with light arrhythmia may re-test.
It is inappropriate for people with serious arrhythmia
to use this blood pressure monitor.
The blood pressure monitor did not inflate
sufficiently.
The cuff was not properly wrapped or the
rubber tube was bent or pressed.
Review the cuff wrapping and testing sections of the
instructions and re-test.
LCD Display
shows abnormal
result
The cuff position was not correct or it was not
properly tightened
Apply the cuff correctly and try again
Body posture was not correct during testing
Review the body posture and testing sections of the
instructions and re-test.
Speaking, arm or body movement, angry,
excited or nervous during testing
Re-test when calm and without speaking or moving
during the test
No response when
you press button
or load battery
Incorrect operation, or strong electromagnetic
interference
Take out batteries for five minutes, and then reinstall
all batteries.
MAINTENANCE
1. Do not drop this monitor or subject it to strong impact.
2. Avoid high temperature and polarization. Do not immerse the monitor in water as this will result in damage to the
monitor.
3. If this monitor is stored near freezing, allow it to acclimate to room temperature before use.
4. Do not attempt to disassemble this monitor.
5. If you do not use the monitor for a long time, please remove the batteries.
•
When the cuff is dirty, detach it from the equipment, wash the cuff by hand with proper detergent and rinse in
cold water. Dry in air. Never iron it.
EXPLANATION OF SYMBOLS ON UNIT
•
Symbol for “Caution Consult Accompanying Documents”
Symbol for “Type B Applied Parts”
CE is a kind of marking of EU safe attestation. This product matches MDD93/42/EEC
Safety precautions:
To reduce risk of electric shock, this product should ONLY be opened by an authorized
technician when service is required. Disconnect the product from mains and other equipment if a
problem should occur. Do not expose the product to water or moisture.
Maintenance:
Clean only with a dry cloth. Do not use cleaning solvents or abrasives.
Warranty:
No guarantee or liability can be accepted for any changes and modifications of the product or damage caused due to
incorrect use of this product.
RISK OF ELECTRIC SHOCK
DO NOT OPEN
CAUTION