5.0 Calibration of the TDX-Series Unit
5.0 Calibration of the TDX-Series Unit
5.1 Checking Channel Response to Gas
The TDX system is supplied precalibrated. After initial installation, the sensor should be checked with calibration/test
gas to confirm gas response. This test should be performed periodically (once every 3 – 4 months).
To check if the sensor requires calibration, introduce an appropriate calibration gas (gas of known concentration) to the
sensor. If there is a discrepancy of more than 10% between the voltage reading on the transmitter and the known
concentration of the gas, then the transmitter should be recalibrated.
Figure 3: Calibration Diagram
TDX-Transmitter
TDX-Sensor
Calibration
Adapter
Flexible
Tubing
Apply Calibration Gas
- Formaldemeter htV (14 pages)
- PPM Formaldemeter™ htV-m (19 pages)
- PGD2 (34 pages)
- PGD2Manual.pdf (28 pages)
- RECON/4 (10 pages)
- RECON-4 (17 pages)
- RECON-IS (15 pages)
- RECON/B SERIES (17 pages)
- RECON Series (16 pages)
- OMNI-4000 (72 pages)
- QUADRANT (26 pages)
- SMARTLOGGER (19 pages)
- SPECTRUM (32 pages)
- SPECTRUM CO-RAL (18 pages)
- SPECTRUM ON-LINE (30 pages)
- SPECTRUM-RAL (15 pages)
- SPECTRUM-RAL-DC (17 pages)
- SPECTRUM SP (20 pages)
- TARGET (36 pages)
- TX-2000 (24 pages)
- AM-5150 (23 pages)
- AM-5175 (21 pages)
- ENG – 97D STAND-ALONE (12 pages)
- GSM-60 (39 pages)
- ISA-60M with MRI-5175 (28 pages)
- MRI-5175 (2 pages)
- MEDAIR 2200 (40 pages)
- PROAIR 2200 (40 pages)
- CP-60 (23 pages)
- EX-5100 (18 pages)
- EX-5175-EC (16 pages)
- ISA-200-RAL (O) (24 pages)
- ISA-40 (19 pages)
- ISA-40M (18 pages)
- ISA-44-2OD (32 pages)
- ISA-44-RALE-OD (38 pages)
- ISA-44-RAL-OD (28 pages)
- ISA-M (15 pages)
- ISA-RAL-M (22 pages)
- MedAir 2000 (30 pages)
- CD-1300-ST (13 pages)
- EX-5120 (18 pages)
- EX-5130 (16 pages)
- EX-5150-MOS (19 pages)