Operational and technical specifications, M6 m50 – VICI M50 User Manual
Page 25

21
Physical Specifications
Dimensions
See Figure 2
See Figure 2
Weight: pump
520 g
540 g
motor and gear assy.
560 g
650 g
micro-electric controller
300 g
300 g
power supply
300 g
300 g
Resolution
MSEL=256
MSEL=256
Minimum flow
5 nL/min
1 µL/min
Maximum continuous flow
5 mL/min
25 mL/min
Maximum intermittent flow*
10 mL/min
35 mL/min
Maximum back pressure
100 psi
100 psi
Volume accuracy (%error)
±0.5% at 1.25 mL
±0.5% at 8 mL
±0.5% at 125 µL
±0.5% at 800 µL
Volume precision (%CV)
<0.1% at 1.25 mL
<0.1% at 8 mL
<0.5% at 125 µL
<0.5% at 800 µL
Fittings
1/16" Valco detail (10-32) 1/8" Cheminert (1/4-28)
Fluid contact
PTFE, Sapphire, PAEK,
PTFE, Ceramic,
Viton
®
, and Valcon P
Valcon H2
Planetary Gear Assembly
Ratio
1:4.75
1:9.88
Motor
Type
2 Phase Bipolar (1.8° per step)
Environmental
Operating Temperature
0 - 50°C
Operating Humidity
< 90% non-condensing
Storage Temperature
-20°C 70°C (dry)
Communications (COM1 or COM2)
Type
RS-232 standard, RS-485 (RS-422) optional
Baud rate
Default 9600 bps, Full range 4800 - 38,800 bps
Parity
None
Data bits per character
8
Stop bits
1
Error checking
None
Mode
ASCII
Power requirement
Rest current draw
0 mA
Operational current draw
Typical 800 mA, 1600 mA at acceleration
Voltage
Typical 24 VDC, input voltage 12 - 48 VDC
Both Models (M6 and M50)
M6
M50
* High speed continuous flow or the use of chemicals incompatible with the wetted surfaces
*
(refer to page 17) will reduce pump lifetime.
Operational and Technical Specifications