Installing in a vacuum system
➨
Assemble the cold trap at a stand mounting.
➨
Connect the cold trap to the vacuum pump and the recipi-
ent.
☞
Prevent mechanical stress at the connection due to tensile
force or vibrations.
➨
Close the condensate drain screw.
During operation
➨
Fill coolant into the coolant chamber.
☞
Check coolant level at appropriate intervals.
➨
Switch on the vacuum pump immediately after filling in the coolant.
☞
The coolant consumption is increased if the condensate chamber is not evacu-
ated
☞
The lifetime of the coolant may be restricted due to icing in case of high operating
pressures.
➨
Vent the cold trap for de-icing or draining of condensate.
➨
Drain condensate via the condensate drain screw. Vent the system prior to open-
ing the drain screw!
☞
Wait until the condensate has liquidised, if necessary.
☞
Ensure that condensate drain tubing and collecting equipment are chemically re-
sistant against the condensate.
☞
Remove the cold trap from the vacuum installation to drain the coolant if neces-
sary.
Use and operation
Maintenance and cleaning
☞
The cold trap is maintenance-free.
☞
Clean the cold trap, coolant chamber and the condensate chamber using water
or solvent. Ensure compatibility of the wetted parts!
☞
Comply with applicable regulations when disposing of solvents.