beautypg.com

VACUUBRAND PC 3001 VARIOpro emission condenser Peltronic User Manual

Page 97

background image

page 97 of 120

Fault

Possible cause

Remedy

❑ ”Vac control” func-

tion: Control stops,

”arrow up” is flash-

ing.

➨ Preset maximum pres-

sure exceeded?

✔ Confirm by pressing

START/STOP key.

Change maximum pres-

sure value if necessary.

❑ ”Pump down” func-

tion: Control stops,

”arrow down” is

flashing.

➨ Pressure below preset

minimum pressure?

✔ Confirm by pressing

START/STOP key.

Change minimum pres-

sure value if necessary.

❑ No function is dis-

played. No menu.

➨ No controllable device

(valve, VARIO pump,

VMS) connected?

✔ Connect devices or use

controller as measuring

device.

❑ Controller does

not respond when

pressing keys (ex-

cept ON/OFF).

PC symbol is dis-

played.

➨ Controller in remote

mode?

✔ Control CVC 3000 via in-

terface or switch off re-

mote mode.

❑ Controller does

not respond when

operating any keys.

No change after

switching off/on.

✔ Contact local distributor.

❑ Pump does not

start or stops im-

mediately. Warning

triangle and pump

symbol are flash-

ing.

➨ Pump has been ex-

posed to condensate?

✔ Allow pump to run for

some minutes at maxi-

mum speed with atmo-

spheric pressure at the

inlet.

➨ Overpressure in outlet

line?

✔ Remove blockage in line,

open valve.

❑ Pump does not

achieve its ultimate

vacuum or usual

pumping speed.

➨ Centring ring at small

flange connection not

correctly positioned, or

leak in the pipeline or

vacuum system?

✔ Check pump directly, then

check connection, pipe-

line and vacuum system if

necessary.

➨ Wrong setting at con-

troller?

✔ Select function ”Pump

down” with speed set to

”HI” and check again.

➨ Long, narrow vacuum

line?

✔ Use lines with larger di-

ameter, length as short as

possible.

➨ Pump has been ex-

posed to condensate?

✔ Allow pump to run for

some minutes with atmo-

spheric pressure at the

inlet to purge.

This manual is related to the following products: