Use and operation – VACUUBRAND BVC 21 User Manual
Page 12
Documents are only to be used and distributed completely and unchanged. It is strictly the users´ responsibility to check
carefully the validity of this document with respect to his product. Manual-no.: 999054 / 15/12/2009
page 12 of 27
Use and operation
Installation:
☞
Max. ambient temperature: 40 °C
➨
Remove the protective foils from the fittings at the flask head and the protective
covers from the inlet of the diaphragm pump or the solenoid operated valve.
➨
Assemble the connecting tube with filter between the bottle head and the inlet of the
diaphragm pump (BVC 21) or the inlet of the solenoid operated valve (BVC 01).
➨
Connect the aspiration tube of the VacuuHandControl VHC to the coupling at the
bottle head.
➨
Connect the female connector of the mains cable to the pressure control device.
☞
Make sure ventilation is adequate if pump is installed in a housing or if ambient
temperature is elevated. Keep a distance of min. 20 cm between fans and ambient
parts.
☞
If pump is installed in altitudes of more than 1000 m above mean sea level check
compatibility with applicable safety requirements, e. g. IEC 60034 (motor may
overheat due to insufficient cooling).
☞
Before switching on the pump check that mains voltage and current conform
with the equipment!
Pump with dual-voltage motor:
☞
Check that the voltage changeover switch underside the terminal box is posi-
tioned correctly.
Attention: If the pump is switched on with the voltage not correctly selected,
the motor may be damaged! Pay attention to the warning label on the on/off
switch and remove after having set the voltage changeover switch to the correct
position!
Selecting the supply voltage:
Attention:
Before starting the pump, check that
the voltage selection switch under-
side the terminal box is correctly
positioned.
Change the selection at the voltage
changeover switch only if the pump
is separated from mains.
Voltage selection switch:
Use a screw driver to adjust the supply
voltage on the voltage changeover
switch at the terminal box of the pump:
”115” corresponds to 100-120V and
”230” corresponds to 200-230V.
During operation
:
The pump achieves its pumping speed, ultimate total vacuum and vapour pumping
rate only at operating temperature (after approx. 15 minutes).
☞
Prevent internal condensation, transfer of liquids or dust. The diaphragm and valves
will be damaged, if liquids are transfered through the pump in significant amounts.