SHEL LAB SSI5R-HS and SSI5R-HS-2 Shaking Incubator Operational Manual

12

Interior Accessory Outlet

This unit features an accessory outlet to provide power
for equipment such as magnetic stirrers, rockers, etc.
The weight of this equipment should not exceed 25
pounds (55.125 kg) per shelf. This equipment may
provide additional heat that could affect the temperature
range of this incubator. We recommend testing the
incubator and any accessory equipment to ensure that
the desired operating conditions can be met.

Caution: This

incubator

operates

at

conditions that might damage
accessory equipment. Verify
that your accessory equipment is
capable of operating under the
same

conditions

as

the

incubator.

The outlet is located inside the incubator in the upper
right rear of the chamber. The voltage available at the
accessory outlet is the SAME as the voltage supplied to
the incubator. For example, a 120-vac incubator will
have 120 vac at the accessory outlet, and a 240-vac
incubator will have 240 vac at the accessory outlet. DO
NOT exceed 500 va rating of the accessory outlet.