beautypg.com

Other – Shellab SMI6E User Manual

Page 10

background image

10

OTHER

Wall fuse/circuit is blown

1/ check wall power source.
2/ compare current draw and compare to specs on data plate.
3/ see what other loads are on the wall circuit.

Unit will not turn on

1/ check wall power source
2/ check circuit breaker for power outlet
3/ check all wiring connections, esp. around the on/off switch

Unit is smoking

– out of box

This is not an uncommon occurrence when the protective coatings heat
up during initial operation. Put unit under vent and run at full power for
one hour until the smoke dissipates.

Contamination in chamber

1/ see cleaning procedure in section 6.1
2/ develop and follow standard operating procedure for specific
application; include definition of cleaning technique and maintenance
schedule.

If following these troubleshooting suggestions does not solve the problem, call Customer Service for
assistance. Please see the manual cover for information on where to contact Customer Service.

This manual is related to the following products: