Arge, Perture, Alorimeter – Scientech Large Aperture Calorimeter User Manual
Page 3: Pecifications, Ertification, Nvironmental, Equirements

3
L
ARGE
A
PERTURE
C
ALORIMETER
S
PECIFICATIONS
:
Model
360401
380401
380402
384UV5
360801
380801
380802
384UV5
Type of Absorber
Surface
Volume
Volume
Volume
Surface
Volume*
Volume
Volume
Aperture Diameter
100mm
200mm
Minimum Beam Diameter
Spectral Response
.25 - 35µm .266 - 1.2µm
9 -11µm
.193 - .36µm .25 - 35µm .266 - 1.2µm
9 -11µm
.193 - .36µm
Maximum Average Power
50W with full illumination of absorbing surface
100W with full illumination of absorbing surface
Minimum Average Power
150mW
700mW
Noise Level
1.5mJ - mW
7mJ - mW
Maximum Power Density
200W/cm²
See Note 1
4W/cm²
Note 2
200W/cm²
See Note 3
4W/cm²
Note 2
Maximum Peak Power Density
1MW/cm²
See Note 4 100MW/cm² See Note 5
1MW/cm²
See Note 6 100MW/cm² See Note 5
Maximum Single Pulse Energy
Maximum Energy Density
Note 7
Note 8
4J/cm
2
Note 9
Note 7
Note 10
4J/cm
2
Note 9
Precision
< 1%
Accuracy
5%
Response Time
5 sec when connected to a Scientech Indicator in Watts Mode
Dimensions DxL - inches/cm
6.00 x 8.00/15.24 x 20.32
9.00 x 10.00/22.86 x 25.40
Weight - pounds/kgs
6/2.72
16.27/7.26
Indicator Compatibility
H410, H410D, S310, S310D
* This is a segmented absorber
Note 1: 380401
27W/cm
2
@ 1064 nm, 21W/cm
2
@ 532 nm, 7.7W/cm
2
@ 355 nm, 158mW/cm
2
@ 266nm
Note 2: 384UV5, 388UV5
50W/cm
2
@ 355nm
Note 3: 380801
13.5W/cm
2
@ 1064 nm, 10.5W/cm
2
@ 532 nm, 3.85W/cm
2
@ 355 nm, 79mW/cm
2
@ 266nm
Note 4: 380401
90GW/cm
2
@ 1064 nm, 71GW/cm
2
@ 532 nm, 27GW/cm
2
@ 355 nm, 530MW/cm
2
@ 266nm
Note 5: 384UV5, 388UV5
Repetitive pulses: 101MW/cm
2
@ 355nm
Single pulses: 3.5GW/cm
2
@ 355nm
Note 6: 380801
45GW/cm
2
@ 1064 nm, 35.5GW/cm
2
@ 532 nm, 13.5GW/cm
2
@ 355 nm, 265MW/cm
2
@ 266nm
Note 7: 360401, 360801
Max J/cm
2
= 1000 x (pulse width)
1/2
to a maximum of 200J/cm
2
Note 8: 380401
Repetitive pulses:
3.7J/cm
2
@ 1064nm, 2.9J/cm
2
@ 532nm, 1J/cm
2
@ 355nm, 20mJ/cm
2
@ 266nm
Single pulses:
7J/cm
2
@ 1064nm, 5.6J/cm
2
@ 532nm, 2.1J/cm
2
@ 355nm, 41mJ/cm
2
@ 266nm
Note 9: 384UV5, 388UV5
Repetitive pulses: 1.1J/cm
2
@ 355nm
Single pulses: 40J/cm2 @ 355nm
Note 10: 38-0801
Repetitive pulses:
1.85J/cm
2
@ 1064nm, 1.45J/cm
2
@ 532nm, 0.5J/cm
2
@ 355nm, 10mJ/cm
2
@ 266nm
Single pulses:
3.5J/cm
2
@ 1064nm, 2.8J/cm
2
@ 532nm, 1.05J/cm
2
@ 355nm, 20.5mJ/cm
2
@ 266nm
150J
300J
5cm
7.5cm
CE
M
ARK
C
ERTIFICATION
:
All of the calorimeters listed in this manual have been certified for the European CE mark.
E
NVIRONMENTAL
R
EQUIREMENTS
:
This product is intended for indoor use at altitudes up to 2000 meters, Pollution Degree 2 in accordance with
IEC 664 and transient overvoltages according to Installation Categories (Overvoltage Categories) II. Note that
each of the above detectors will not pass the IEC 801 Publication, Part 3, Radiated Electromagnetic Field
Requirements. The system, meter and detector, is designed to measure radiation within the test's radiation band.
The detectors were held outside the radiated electromagnetic field during this test. It is up to the user to be
aware of RF fields present during measurements and their effects if any on those measurements.